AN5800/40 eccentric module
AN5800/40 eccentric module
Classification:
Product Description
Technical data
1. Sensor input:
Proximity Sensors:
Sensitivity: 8V/mm, ±1mm range;
Output voltage range: -2~-18VDC;
Initial installation zero voltage: -10VDC.
Note: An eddy-current sensor needs to be installed on each of the key phase and eccentricity, which outputs the key signal and eccentric signal to the eccentric channel and the key phase channel of the module respectively.
2. Measuring range:
Eccentric range: 0~999μm (peak-to-peak: Pk-Pk) Accuracy: ±0.5%±1μm
Keyphasor output: TTL level signal
3. Module power supply:
DC power supply: 24VDC(18~36V)
Module power consumption: less than 5W
4. Standard signal output:
Standard current: 4~20mA (corresponding to the upper limit of full scale)
Active: Drive load: ≤500Ω, effective transmission distance: ≤ 300 m (1000 ft).
5. Switching output:
Two-stage threshold (Alert, Danger) programmable setting and normally open and normally closed dry contact output, the full scale range can be set, and the error is ≤0.1%;
Insulation resistance: ≥1000MΩ;
Contact capacity: 2A/30VDC, 1A/220VAC; Double Pole Double Throw (DPDT);
Alarm delay: 0.1 to 60S, default value: 0.5S;
Alarm Latching: The default is non-latching.
Key phase channel: Only Not-ok relay contact output
6. Physical characteristics:
Front Plate Size: (L×W×H) 248mm×40mm×262mm 9.8in× 1.5in×10.3in
Weight: 600g
Rear Insert Board Size: (L×W×H) 128 mm× 40 mm× 262 mm 5 in× 1.5 in× 10.3 in
Weight: 300g
7. Environmental restrictions:
Operating temperature: -30°C~+65°C(-22oF~+150 oF)
Storage temperature: -40°C~+85°C(-40oF~+180 oF)
Relative humidity: 0~95% non-condensing
Refer to the implementation standard
American Petroleum Institute API 670 standard;
Safety Certification: C E;
Anti-electromagnetic interference: EMC\ LVD;
Refer to EN55011\EN 61010\EN50081\EN50082.
Product selection
AN5800/40 -AXX-BXX-CXX
A: Channel selection
00—Dual channel (key phase, eccentricity)
01—Single channel (only used as a key phase)
B: Sensor selection
00—φ8/φ5 diameter eddy current sensor
C: Relay output selection (eccentric channel)
00—Three DPDT dry contact outputs (Alert, Danger, Not-ok)
01—Two DPDT dry contact outputs (Alert, Danger)
02—Single DPDT Dry Contact Output (Alert)
03—No relay output